Physical Failure Checklist
Here are physical failures that can result in failure to desaturate.
#1: Enthusiastic user cranked the oxygen flow meter too high
The oxygen flow meter should be set to 10 LPM or maybe slightly above. Some new users will open the oxygen flow to maximum and the ball will be at the complete top of the oxygen flow range. This error will also cause the air separator to display an amber light that will never go off.
This pushes so much air through the air separator that both the oxygen concentration in +O2 is low, and the oxygen concentration in the ‑O2 is high. The extra oxygen at the ‑O2 setting often prevents users from desaturating.
This pushes so much air through the air separator that both the oxygen concentration in +O2 is low, and the oxygen concentration in the ‑O2 is high. The extra oxygen at the ‑O2 setting often prevents users from desaturating.
#3: Ripped breather hose
This occurs when the breather hose is pulled so hard the membrane between the ribs rips and allows room air to leak in when the user inhales.
This leak allows room air to dilute both the +O2 and ‑O2 air which inhibits both oxygen saturation and oxygen desaturation.
The best way to test this is to carefully inspect the hose to assure the membrane between the ribs is intact for the entire length of the hose.
Tutorial: Inspect and replace a damaged breather hose
If you doubt your inspection, remove the breather hose from the reservoir and plug one end with your thumb. Air should not escape when you blow air in from the unplugged end.
If air escapes or you find a tear you will need to replace the breather hose.
#4: Poor Mask Seal
- Make sure the mask is sealed properly- suck test. Put the mask on your face and inhale while blocking the upper air inlet port. This should cause the mask to draw to your face.
- Use a mirror to make sure the mask valves are opening and closing correctly when the mask is on.
- Make sure to not touch the silicone valve flaps when putting on the breather hose.
#5: Pulse Oximeter Issues
- Make sure the batteries are not low.
- Make sure the user does not have any type of fingernail polish.
- Make sure they are not grasping the handlebars too tightly
- Change fingers if the reading is slow to respond.
- Make sure they don’t have excessive movement of the finger.
See Also: Troubleshoot Your Pulse Oximeter
#6: Failed Air Separator (LiveO2 AC)
Each air separator has an internal tester to assures air is separated into oxygen rich, +O2, over 75% and oxygen reduced, ‑O2 mixtures, below 16%.
This is the amber light tester assures that oxygen is at least 75% pure. At a 10 LPM flow rate, it also assures that the ‑O2 level is less than 16%.
This test allows LiveO2 users to know their system is always working without need for repeated testing. We test all air separators before shipping, but they can fail over time, and sometimes during transport.
When an air separator fails, the amber light glows like in the image below. If you see this then go to the amber light diagnostic.
#7: Failed Valve
This one is very rare. Every system is pressure tested before shipping.
Switches between +O2 and ‑O2 should cause a distinct and audible latch as the valve locks into the opposite position. This latch mechanism assures three things:
- The mechanical components of the valve are working properly each time the user switches;
- The valve fully switched between ‑O2 and +O2 air to disallow mixing inside the valve;
- The valve is internally sealed.
If you there is no distinct latch. Please contact customer service.
#8: Oxygen to Hypoxic Air Leak
If you are still having problems disconnect the breather hose from the black valve and insert it into the overflow release port, or use the ‑O2 zipper to trap the breather hose into the ‑O2 compartment.
This isolates the air supply to only ‑O2 bypassing the oxygen reservoir and valve. Exercise breathing this air. If you desaturate with this configuration differently than on the ‑O2 setting then oxygen is mixing with the hypoxic air inside the reservoir.
Please contact customer service to arrange repair or replacement of your reservoir.
Please note the hole in the ‑O2 compartment is not a defect. It releases excess air.
All else fails Order a Tester
Contact support and we can provide a separate oxygen tester that will enable you to test the actual output of oxygen from the reservoir at the mask.
We put this last because every air separator has a built-in tester that continuously tests the oxygen level created by the air separator. This built-in tester continuously assures proper oxygen concentration.
Our support department can optionally provide an oxygen tester to confirm the levels are correct as delivered to the mask. Oxygen Test Meter Deposit
Related articles: